Rectangular Chamber With Smooth Round Corners And Jacket Medical Instrument
Sterilizer Machine
Model | MDL-0.24 | MDL-0.36 | MDL-0.6 |
Chamber Size(W*H*D mm) | 600*600*670 | 600*600*1000 | 610*910*1170 |
Overall Size(W*H*D mm) | 1350*1800*1010 | 1350*1800*1340 | 1450*1900*1510 |
Steam(Kg/cycle) | 20 | 25 | 30 |
Water(Kg/cycle) | 30 | 40 | 60 |
Power(KVA) | 2 | 2 | 3 |
Applications | Sterilization of Liquids | ||
Unwrapped instruments and glassware | |||
Wrapped instruments, packs and utensils |
Sterilizers shall be horizontal with vertically opening double
door, horizontal chamber
type (rectangular chamber with smooth round corners and jacket)
with fully automated operating
system. The entire construction including all components shall be
fully compliant with cGMP
requirements
Sterilization Chamber
The chamber shall be rectangular, with smooth and rounded corners.
The chamber shall be designed as per GB150 pressure vessel code.
The chamber shall be made of AISI 316L or 304 with surface roughness less than 0.5µm.
The chamber shall be re-inforced with an AISI 304 jacket.
The sterilizer support frame for the entire structure shall be made of
AISI 304.
The sterilizer shall be able to reach and maintain sterilization
temperature of 121°±1.0°C
Chamber Doors
All Steam sterilizers shall have double door.
The door made of AISI 316L with internal surface roughness less
than 0.5µm.
Sliding door models. Door opening and closing shall be automatic.
The door gaskets made of high temperature resistant silicone rubber
with rounded corners.
Door Safety
The following door safety features provided for operator safety:
Door interlocking to prevent simultaneous opening of both the
doors.
Door process lock to prevent opening of doors when the process is
on.
Sterile side door to open only after sterilization cycle.
Insulation
The sterilization chamber shall be insulated with thick mineral
wool, hold in place by an outer
cover of AISI 304. Chamber insulation shall be in such a way that
the external surface
temperature shall not exceed 50ºC during operation.
Piping
All piping made of AISI 316 with internal surface roughness less
than 0.5µm.
All welds on piping made with orbital TIG method.
All non contact surfaces shall be of AISI 304.
All pipes installed to ensure the possibility to drain them
completely and avoid condensate residuals..
Steam traps of sanitary design shall be provided in the steam lines
for condensate removal.
All the pipe fittings shall be of sanitary Tri-clover type.
Medpharm medical large steam autoclave brochure.pdf
ZHANGJIAGANG MEDPHARM MACHINERY LTD.
MEDPHARM Company team is committed to providing quality sterilization products, service, parts and
maintenance training at significant savings. From customized parts packages to turnkey equipment
acquisition and installation, our innovative sterilization solutions are designed to meet our customers' needs.
We understand that untimely breakdowns are costly and can be detrimental to your everyday operations.
Our service team is compromised of expert technicians that respond quickly to all of your service needs.
We offer on-site service support including inservices, installation and deinstallation of your equipment. From
breakdown repairs to comprehensive maintenance programs, we have an innovative service solution for you.
We will service on-Site Repairs.Our staff is equipped with the resources required to provide you with
on-site, on-time repairs in compliance with manufacturer specifications. We also provide installation,
deinstallation, and inservices on most makes and models of sterilization equipment. Near or far we can
provide you with service support that you need.
Our reconditioning background has taught us of the importance of having quality parts on hand at all times.
We have a large inventory of individual repair parts, as well as preventive maintenance kits, and consumables
in stock and ready to ship today.
With our expert technical support staff, it is not always necessary to have part numbers to place an order.
We will help you identify the quality parts that you need.